🔬 Trace Analysis Lab: Limit of Detection & Quantification (LOD/LOQ)

Interactive 3D analytical-chemistry instrument: a spectrometer measures samples through real baseline noise while building a live calibration curve from added standards. Adjust instrument noise and test-sample concentration, watch the least-squares slope and the blank-replicate standard deviation update in real time, and see the IUPAC limit of detection (LOD = 3.3sigma/S) and limit of quantification (LOQ = 10sigma/S) plotted directly on the calibration graph as the sample signal rises out of the noise.

Chemistry & Materials3DAdvanced60 FPS
Trace Analysis Lab: Limit of Detection & Quantification ↗ Open standalone

A 3D spectrometer measures samples through real baseline noise while building a live calibration curve from standards — watch the IUPAC limit of detection and limit of quantification appear on the graph as the sample signal rises out of the noise.

🔬 What It Demonstrates

Every real instrument has baseline noise, quantified as sigma (σ), the standard deviation of repeated blank measurements. Comparing σ against the calibration curve's least-squares slope (S) gives the IUPAC limit of detection (LOD = 3.3σ/S) and limit of quantification (LOQ = 10σ/S) — the concentrations below which a signal can't be reliably called "present" or "measured," respectively.

🎮 How to Use

Adjust instrument noise and test-sample concentration, then click Add calibration point to record a fresh noisy standard and refit the calibration line. Watch σ, the slope, and LOD/LOQ update live, and see the test sample's diamond marker shift color as it crosses the LOD and LOQ lines on the graph.

💡 Did You Know?

A measurement right at the blank's own average isn't proof nothing is there — it's simply below LOD, meaning any real analyte at that level is statistically indistinguishable from noise, which is why trace-contaminant reports are written as "< LOD" rather than as a false zero.

⚙ Under the hood

Adjust instrument noise and sample concentration to find the Limit of Detection (LOD) and Limit of Quantification (LOQ) on a calibration curve.

Three.jsChemistryAnalytical ChemistryTrace AnalysisChemical MetrologyCalibration CurveLODLOQ