🔬 Trace Analysis Lab: LOD/LOQ

Calibration slope (S): —
Blank SD (σ): —
Blank replicates: 0
LOD = 3.3σ/S: —
LOQ = 10σ/S: —
Sample signal: —
Sample status: —
Standards: 0
FPS: —
Calibration curve — dots are standards, dashed lines mark LOD and LOQ, diamond is the test sample
Drag — rotate · Scroll — zoom

🔬 Trace Analysis Lab: Limit of Detection & Quantification (LOD/LOQ)

A 3D spectrometer measures samples through real baseline noise while building a live calibration curve from standards — watch the IUPAC limit of detection and limit of quantification appear on the graph as the sample signal rises out of the noise.

🔬 What It Demonstrates

Every real instrument has baseline noise, quantified as sigma (σ), the standard deviation of repeated blank measurements. Comparing σ against the calibration curve's least-squares slope (S) gives the IUPAC limit of detection (LOD = 3.3σ/S) and limit of quantification (LOQ = 10σ/S) — the concentrations below which a signal can't be reliably called "present" or "measured," respectively.

🎮 How to Use

Adjust instrument noise and test-sample concentration, then click Add calibration point to record a fresh noisy standard and refit the calibration line. Watch σ, the slope, and LOD/LOQ update live, and see the test sample's diamond marker shift color as it crosses the LOD and LOQ lines on the graph.

💡 Did You Know?

A measurement right at the blank's own average isn't proof nothing is there — it's simply below LOD, meaning any real analyte at that level is statistically indistinguishable from noise, which is why trace-contaminant reports are written as "< LOD" rather than as a false zero.