Pan-Tompkins QRS Detector: Pipeline Traces, Filter Responses and Detection Scores (2D)
The same detector as four scrolling traces, one per pipeline stage, with a pulsing heart, true and detected beats and the adaptive threshold, above the filter gain curves with the QRS band and the 50 Hz hum marked and bars for sensitivity and positive predictivity.
Lie a Blender-modelled patient on a hospital bed, stick three electrode pads on the chest and watch a bedside monitor run the classic Pan-Tompkins QRS detector over a synthetic 200 Hz ECG. Set the heart rate, beat-to-beat variability, baseline wander, mains hum and muscle noise, change the integration window and the threshold fraction, add premature ventricular beats, and see sensitivity, predictivity and R-wave location error against the known beat times. The filter recursions are checked against their closed-form frequency responses.
The same detector as four scrolling traces, one per pipeline stage, with a pulsing heart, true and detected beats and the adaptive threshold, above the filter gain curves with the QRS band and the 50 Hz hum marked and bars for sensitivity and positive predictivity.
2D · HTML5 Canvas 2D · 60 FPS target · runs fully client-side, no install