A differential scanning calorimeter (DSC) compares the heat flow needed to follow a programmed temperature ramp for a sample and an inert reference. A transition appears as a peak or step in the differential signal.
T(t) = T₀ + βt
For a simple constant-rate scan, the baseline difference is approximated by:
Φbase = β · m · ΔCp
An illustrative Gaussian transition peak is added:
Φpeak(T) = (m · ΔH · β / (√(2π) · σ)) · exp(−(T − Tp)² / (2σ²))
The peak area recovers the transition enthalpy:
ΔH = (1 / m) · ∫ Φpeak(t) dt
Model assumptions: a linear heating ramp, constant sample mass and heat-capacity difference, one symmetric transition peak, no thermal lag, no pan resistance, and ideal baseline subtraction. The displayed curve is an educational model, not instrument calibration data.
Independent area check: calculating…
Readouts and equations are illustrative and scaled for teaching. The mesh is a representative cutaway, not a manufacturing drawing.