Remote centre of motion. In laparoscopic surgery the instrument enters the body through a small port in the abdominal wall. The port is a fixed point the shaft must pass through, so the instrument can only pivot about it, slide in and out, and spin: a remote centre of motion (RCM). If the tip is wanted at q and the port is at P, the shaft direction is u = (q - P)/|q - P| and the wrist hub sits outside the body at H = P - u (Lsh - |q - P|). A three-joint arm (base yaw, shoulder, elbow, link lengths L1 = 40 cm, L2 = 38 cm) reaches H by planar two-link inverse kinematics, cosθ2 = (r² + y² - L1² - L2²)/(2L1L2). If the robot's assumed pivot is wrong by e (a calibration error), the shaft is pushed sideways at the port by e√(1 - ux²) and the wall loads it with about Kw times that displacement (Kw = 0.25 N/mm here, an illustrative value).
Motion scaling and tremor filtering. The surgeon's hand path (minimum-jerk movements between four targets) plus a physiological tremor A sin(2πft) at 6 to 12 Hz is divided by the scaling ratio and passed through a first-order low-pass filter with time constant τ. The filter attenuates a tremor at frequency f by 1/√(1 + (2πfτ)²), but it also delays the intended motion: a ramp of speed v trails by vτ and a step needs τ ln 50 to reach 98 %. The lower graph is this trade-off: tremor error falls and lag error rises with τ, so a best τ exists (the readout shows it).
Checks. The arm is rebuilt by forward kinematics from its own joint angles and the port-to-shaft distance is compared with the closed form; the filter's tremor gain is measured by lock-in detection at 0.1 ms steps; the ramp lag and the 98 % settling time are measured against vτ and τ ln 50; and the full pipeline's RMS tip error is compared with the closed-form sqrt(tremor² + lag²).
Simplifications. A single point-mass tip with an ideal inner position loop (no joint dynamics, backlash, cable stretch, haptics or camera delay); tremor is one sinusoid in a fixed direction; the filter is first order (real systems often use notch or adaptive band-limited filters); the abdominal wall is a linear spring and the shaft is rigid. This is an educational model of the principle, not a medical device description and not affiliated with any robot maker. Numbers are illustrative orders of magnitude.