How do you weigh something you cannot see?
Instead of putting matter on a scale, a quartz crystal microbalance (QCM) measures how added material changes the resonance of a piezoelectric quartz plate. A thin AT-cut crystal is sandwiched between electrodes and driven in thickness-shear vibration. Its surface moves sideways, often by a few nanometres or less. Added rigid surface mass makes the resonance frequency fall. A surrounding liquid does something different: it both loads the surface and drains energy from the oscillation. QUARTZFIELD links these effects through six experiments with a consistent, equation-driven model.
01. Deposit a rigid film
Start with a dry surface. Move the film-thickness control while monitoring Δf. For a thin, laterally uniform, perfectly attached rigid coating, the Sauerbrey equation gives Δfₙ = −2nf₀²σ/Zq, where n is odd overtone number, f₀ is 5 MHz, σ is areal mass in kg/m², and Zq = √(ρqµq) is the quartz acoustic impedance. Here ρq = 2648 kg/m³ and µq = 2.947×10¹⁰ Pa. Hypothesis: doubling film thickness at fixed density doubles the magnitude of Δf. Change the density to test that only mass per area matters in this approximation.
02. Investigate shear harmonics
Switch between n = 1, 3, 5, 7, and 9. The ideal thickness-shear standing displacement across quartz is u(z) ∝ cos(nπz/t). Each higher odd mode has more interior nodal surfaces. The unloaded resonance is nf₀, while the mass-driven shift is n times the fundamental shift. The second graph divides Δf by n: a perfectly rigid Sauerbrey film gives a flat overtone-normalised response. In the 3D instrument, exaggerated transparent layers show the mathematical eigenfunction. The apparent thickness and motion are enlarged for visibility; they are not a scale drawing.
03. Couple the resonator to liquid
Activate liquid contact and vary viscosity η and density ρ. In the ideal Newtonian, no-slip, semi-infinite fluid, an oscillating wall launches a transverse viscous disturbance with 1/e penetration depth δ = √[η/(πfₙρ)]. The hydrodynamic frequency shift is Δfₗ,ₙ = −fₙ³ᐟ²√(ρη/π)/Zq. At this order, the increase in resonance half-bandwidth ΔΓₙ has the same magnitude as −Δfₗ,ₙ. Hypothesis: quadrupling viscosity doubles the liquid frequency shift and bandwidth increase. Probe higher n to see penetration become shallower, even while overall loading increases.
04. Sweep the electrical resonance
A Butterworth–Van Dyke (BVD) circuit represents a quartz mechanical mode as a motional series R–L–C branch in parallel with the static electrode capacitance C₀. Its complex admittance is Y(ω) = jωC₀ + 1/[R + j(ωL − 1/ωC)]. The circuit controls set a reference motional capacitance, shunt capacitance, and an air quality factor. The observed loaded resonance and damping determine L and R consistently. Shift the drive frequency through the measured peak and inspect both magnitude and phase. In this teaching model, higher-overtone capacitance and air-Q scaling are explicitly phenomenological.
05. Release the oscillation
In ringdown mode, a resonator is imagined to be excited and then released at t = 0. The physically derived amplitude envelope is A(t)/A(0) = exp(−t/τ), where τ = Q/(πf). The stored energy falls twice as quickly in its exponent: E(t)/E(0) = exp(−2t/τ). Compare dry and wet configurations. The 3D motion is displayed stroboscopically, because 5–45 MHz vibrations cannot be individually shown on a 60 Hz screen. The interface magnifies the millisecond time window: at 1× playback, one simulated millisecond occupies one real second. Choose play, pause, reset, and speed to explore the physical envelope; the quartz vibration itself is shown stroboscopically.
06. Recover unknown coating mass
The last experiment inverts the five overtone measurements. A least-squares fit estimates the single unknown rigid areal mass σ from Δfₙ = −knσ, k = 2f₀²/Zq. When liquid is present, subtract its known blank contribution before fitting. Compare the naive and corrected estimates in the graphs. A wet resonator without a liquid blank would wrongly attribute viscous loading to a rigid coating. The exact-model corrected residual is zero within floating-point precision; in real instruments, noise, surface heterogeneity and soft viscoelastic layers complicate inversion.
Operate the laboratory
Choose an experiment, then adjust the labelled controls. Drag the turquoise 3D probe over the electrode; drag elsewhere to orbit, scroll or pinch to zoom, and Shift-drag or right-drag to pan. Record multiple conditions in the notebook; CSV contains model parameters and computed readings, while PNG saves the live instrument. Hover graphs for coordinates; in sweep and ringdown modes clicking the first graph positions the frequency or time cursor. Use the challenges for numerical reasoning and live configuration checks.
Limits, language and applications
The equations assume a linear AT-cut shear oscillator, small perturbations, one loaded liquid face, and a rigid uniform film. They omit detailed lateral mode trapping, electrode stress, finite liquid depth, temperature drift, surface slip, viscoelastic film dispersion, acoustic leakage, and instrumental noise. Rendering amplitudes and dimensions are explicitly exaggerated. The movable probe samples a simplified Gaussian lateral vibration envelope; its local amplitude readout is not an independent resonance-frequency measurement. Useful applications include thin-film process monitoring, adsorption kinetics, affinity biosensing, fluid-property studies, and electrochemical interface research. Glossary: overtone, an odd shear resonance multiple; Γ, resonance half-bandwidth; Q, stored-to-dissipated energy measure; dissipation D, 1/Q; areal mass, mass per unit sensor area. Further questions: how would a slipping film break Sauerbrey behaviour, and how could a liquid reference reduce systematic error?