The science
A silicon microcantilever is a driven, damped oscillator. For a clamped-free Euler-Bernoulli beam the first mode has f0 = (β²/2π) √(EI / ρA L&sup4;) with β = 1.8751 (root of cos β cosh β = -1), i.e. f0 = 0.1615 (t/L²) √(E/ρ).
A small added mass dm at position x lowers the resonance by Rayleigh's principle: f = f0 / √(1 + dm φ(x)² / meff) with the mode shape φ(1) = 1 and modal mass meff = m ∫φ²dξ = 0.25 m. Sensitivity therefore goes as φ²: largest at the free end, almost zero near the clamp.
The limit of detection is set by thermomechanical noise: δf = (f0 / 2Q)(xth / a) √(B Q / f0), xth = √(kBT / k), and the minimum detectable mass is δm = 2 meff δf / f0 (Ekinci, Yang and Roukes, 2004).
Simplifications: single (first) mode, point-mass analytes with no added stiffness or surface stress, constant Q (no damping change from the analyte), Gaussian frequency noise, vacuum-like Lorentzian response, silicon with E = 169 GPa. The plotted vibration amplitude is exaggerated and slowed for visibility. Real liquid-phase sensors lose Q to fluid damping.