Motion held
in a field.
Five experimental stations. One charged ion. Discover how crossed electrical and magnetic forces create a precision instrument for the invisible.
A single ion, three motions.
Watch the combined fast cyclotron orbit, slow magnetron drift and axial oscillation. Drag the glowing ion to set an initial position, or adjust the electrode separation directly.
The architecture
of confinement.
The equations here are the equations running the experiment. Every number on the instrument comes from the configured fields, chosen isotope, or integrated trajectory.
8 SECTIONS
CLASSICAL MODEL
01 What is a Penning trap?+
A Penning trap confines a charged particle without touching it. A static magnetic field along the trap axis bends transverse motion through the Lorentz force, while a static electric quadrupole creates an axial potential well. Neither the magnetic field nor the electrostatic configuration by itself creates three-dimensional stable confinement. Together they can hold an ion for precise spectroscopy and mass measurements. In this laboratory the ion is singly positively charged, the magnetic field is directed along +z, and we ignore any space-charge effects from other ions.
02 How to operate the apparatus+
Begin with the Balanced orbit preset. Drag empty space in the 3D view to orbit the camera; use Shift-drag to pan and the wheel or two-finger pinch to zoom. Drag the glowing ion across its current horizontal plane to reposition it, or move it with the arrow keys after focusing the 3D scene. The brass d-caliper changes the characteristic electrode scale; Shift+arrow keys are its keyboard alternative. The control sliders adjust B, U, d, initial amplitudes, and the optional RF field. Pause, reset or vary the simulation playback speed. The traces record actual calculated positions, rather than decorative animation. Hit Record observation to add a row to your notebook; export all rows as CSV.
03 Governing equations and units+
The ideal electrostatic potential is Φ = U(2z² − x² − y²)/(2d²), where U is in volts and d, x, y, z are in metres. It obeys Laplace’s equation. The electric field is E = −∇Φ = (U/d²)(x, y, −2z). A positive ion follows m d²r/dt² = q(E + v × B), with B = Bẑ. For a vacuum model, q = +e = 1.602176634 × 10⁻¹⁹ C and masses are converted from atomic mass units. The computational clock uses τ = ωc t, where ωc = qB/m. Positions are stored in millimetres, but each force calculation has the corresponding unit conversion. The classical fourth-order Runge–Kutta method integrates the state at fixed steps of Δτ = 0.04, independent of the display frame rate.
04 The three normal motions+
In stable operation the axial angular frequency is ωz = √(2qU/md²). The free cyclotron frequency is ωc = qB/m. Two radial eigenfrequencies are ω± = [ωc ± √(ωc² − 2ωz²)]/2. The faster modified cyclotron orbit and slower magnetron drift rotate in the same transverse direction for the chosen positive-charge convention. Their sum equals ωc. The simulator launches a superposition with selectable fast and slow radii and an independent axial amplitude. The Eigenmode Studio shows a discrete spectral analysis of the exact ideal motion, with a Hann window; finite frequency resolution can broaden or merge peaks. The displayed theoretical mode markers come from the actual field parameters.
05 Five guided investigations+
Orbit Chamber: vary B and U, identify the fast spiral superimposed on a slow guide-centre drift, and inspect position traces. Eigenmode Studio: reduce the fast amplitude, then the slow amplitude, to isolate peaks and compare their frequencies. Stability Atlas: click the B–U diagram; find where the discriminant ωc² − 2ωz² changes sign, then cross the boundary and observe loss of radial confinement. RF Resonance: inject Eₓ(t) = E₀ cos(ωRF t), tune near ω₊, and watch the time-dependent orbital response. Mass Metrology: switch between calcium isotopes or heavier ions while keeping the trap fields fixed, and infer mass from m = qB/[2π(f₊ + f₋)]. Record paired measurements and compare their ratios.
06 Measurements and predicted outcomes+
Hypothesis A: increasing B at fixed U raises the free cyclotron frequency while making radial confinement more robust. Hypothesis B: raising U increases axial frequency, but too much U makes the radial eigenfrequencies non-real. Hypothesis C: heavier ions at the same B have a lower free cyclotron frequency. Investigate each by altering one parameter at a time and saving notebook rows. Without RF the conserved total is H = ½m|v|² + qΦ. Monitor the fractional energy drift to test numerical accuracy; the magnetic force must never change the total energy. With RF switched on, energy need not be conserved because the oscillating electric field can do work.
07 Real-world applications and cautions+
Penning traps support high-precision atomic mass measurements, fundamental-constant tests, quantum-state studies, and charged-particle spectroscopy. Real instruments use shaped electrodes, vacuum vessels, magnetic coils or superconducting magnets, electrical pickup circuits, and techniques for cooling ions. This virtual apparatus is a schematic illustration of that equipment. It imposes the ideal analytic quadrupole field instead of solving the field from the drawn electrode surfaces. It omits relativistic effects, damping, gas collisions, field imperfections, quantum transitions, image charges, detector noise, and RF circuit back-action. Unstable trajectories may leave the displayed chamber; the mathematical state remains calculated without simulated wall impacts.
08 Glossary and open questions+
Cyclotron: circular motion caused by the magnetic Lorentz force. Magnetron: slow radial drift associated with combined electric and magnetic fields. Quadrupole: an electric potential with second-order spatial variation. Eigenfrequency: a natural normal-mode oscillation rate. Stability discriminant: the expression ωc² − 2ωz² governing real radial modes. RF drive: a time-varying electric excitation. Windowing: weighting a finite measurement sequence to reduce spectral leakage. Questions to explore: why does the electrostatic potential destabilise radial motion when it stabilises axial motion? How could you distinguish two isotopes from their frequency ratio? What changes when RF work is included in the energy budget?