Educational physics of sterile-instrument processing, not a validated sterilisation procedure. Real cycles are specified, validated and monitored under national standards and manufacturer instructions. Nothing here is medical or infection-control advice.
Heat reaches steel by conduction
Each item is a flat steel plate of half-thickness L heated from both faces. Inside, rho c dT/dt = k d2T/dx2 (steel: k = 15 W/m/K, rho = 7900 kg/m3, c = 500 J/kg/K, typical values). At the faces, steam condenses with film coefficient h, optionally in series with a textile wrap of resistance R_w, so the surface conductance is U = 1/(1/h + R_w) and heat flow is U (T_steam - T_surface). Residual air hinders condensation; the page reduces h by 1/(1 + 0.1 x air%), an illustrative rule.
Biot number and the lumped model
Bi = U L / k compares surface resistance with internal resistance. For Bi < 0.1 the item is nearly isothermal and T = T_s - (T_s - T_0) e^(-t/tau) with tau = rho c L / U. For thick items (the steel block) the centre lags the surface; the exact solution for a step in steam temperature is the Fourier series theta = sum C_n e^(-lambda_n^2 Fo) with lambda_n tan lambda_n = Bi, C_n = 4 sin lambda_n / (2 lambda_n + sin 2 lambda_n) and Fo = alpha t / L^2. The simulator uses a finite-volume Crank-Nicolson solver (30 cells, Rannacher start) and checks it against this series.
Steam pressure
Saturated steam temperature and pressure are tied together; the gauge uses the Antoine fit log10 P(mmHg) = 8.14019 - 1810.94/(244.485 + T), close to steam tables (about 205 kPa absolute at 121 degrees C, 304 kPa at 134 degrees C). The gauge shows pressure above atmosphere.
Lethality, F0 and log reduction
Thermal kill is modelled as first-order with a D-value (time for a factor-10 reduction) and a z-value (temperature change that changes D by a factor 10). With z = 10 K the equivalent minutes at 121.1 degrees C are F0 = integral of 10^((T - 121.1)/10) dt, taken at the coldest point of each item, and the log reduction is F0 / D. The target F0, D-value and the 1.5 min used here are illustrative; actual requirements come from standards for the specific load. Hold needed is estimated by running the program with a long hold and finding when each item first reaches the target F0.
Simplifications
One-dimensional plates (no hinges, lumens or porous loads), constant material properties, a prescribed linear chamber temperature program instead of real air-removal pulses, no condensate heating of the pack, no spatial steam variation. This 2D view shows the temperature field the same solver computes through the thickness of each item (the strips are drawn at a convenient width each, not to a common scale); the physics runs in JavaScript.