Through-port spectrum
ALL-PASS RING / LIVE RESPONSELive instrument readout
SI / DERIVEDThe science · model and assumptions
Binding: θ = c / (Kᴅ + c). The adsorbed layer shifts effective index by Δnₑff = Δnₛₐₜ θ.
Round-trip resonance: mλ = 2πR nₑff. The ring radius is calibrated so mode 176 is at 1550 nm before binding.
Coupled-mode transmission: T = ( (γᵢ−γₑ)² + Δω² ) / ( (γᵢ+γₑ)² + Δω² ), with γ = ω₀/(2Q) and Qₗ⁻¹ = Qᵢ⁻¹ + Qₑ⁻¹.
Illustrative design: the concentration-to-index overlap and exponential gap-to-coupling relation are representative fitted parameters, not a calibration for a real device. Constant group index, single-mode operation, uniform binding and a single resonance are assumed. Values are educational, not measured biological data.