A dispersive single-beam instrument selects a narrow wavelength band, sends it through a sample and measures the transmitted optical power. A blank is measured separately to establish the incident reference.
A = ε c l
Beer–Lambert law: absorbance A is the product of molar absorptivity ε (L·mol⁻¹·cm⁻¹), concentration c (mol·L⁻¹) and path length l (cm).
T = I / I₀ = 10⁻ᴬ
Transmittance T is the transmitted intensity divided by the reference intensity. The detector current is modelled as Id = R P₀ T, where R is responsivity and P₀ is incident optical power.
m λ = d (sin α + sin β)
For a diffraction grating, order m, groove spacing d, incidence angle α and diffraction angle β determine the selected wavelength. The model shows the grating and slit arrangement schematically.
Independent numerical check: the displayed absorbance is recovered from the computed transmittance using −log₁₀(T). This checks the forward and inverse Beer–Lambert calculations.
Simplifications: ε is held constant as wavelength changes; scattering, stray light, spectral bandwidth, detector dark current, lamp drift and optical losses are omitted. The current is a scaled illustrative estimate, not a performance specification. The instrument geometry is simplified and not dimensioned for manufacture.
Illustrative design, not affiliated with any real institution or manufacturer.