An Abbe refractometer estimates a liquid’s refractive index by observing the boundary between illuminated and dark fields at the critical angle. A sample film sits between a measuring prism and a hinged cover prism.
nsample = nprism · sin(θc)
For the illustrative values used here, the prism index is 1.750 and the sample index is adjustable from 1.30 to 1.60. The critical angle is calculated as θc = asin(nsample/nprism).
nD20 ≈ nDT + α(T − 20 °C)
The temperature correction uses an illustrative coefficient α = 4.5 × 10−4 K−1. Actual coefficients depend on the liquid and wavelength; a real instrument requires calibration and a defined reference wavelength.
Simplifications: monochromatic light is assumed; prism index is treated as constant; dispersion and sample non-uniformity are omitted; the displayed geometry is scaled for inspection rather than manufacture.
Independent check: the page reconstructs the sample index from the calculated critical angle using Snell’s relation and reports the absolute residual.