The science
The airway model
The lung is a symmetric Weibel-type tree of 24 generations: generation 0 is the trachea (1.8 cm wide, 12 cm long), each generation splits into two, down to the respiratory airways in generations 17-23. The page uses rounded Weibel model A diameters and lengths. A breath at mean flow Q sends Q / 2^z through each tube of generation z. The 3D and 2D trees are condensed to 8 display levels that each cover several generations.
Diffusion of a nanoparticle
Air molecules kick small particles around. The diffusion coefficient is the Stokes-Einstein value with the Cunningham slip correction: D = k_B T Cc / (3 pi eta d), Cc = 1 + Kn (1.257 + 0.4 exp(-1.1/Kn)), Kn = 2 lambda / d, with air mean free path lambda about 70 nm at body temperature. A 10 nm particle diffuses about 100 times faster than a 100 nm one.
Deposition in one airway
In laminar flow through a tube of length L, the fraction of particles that reach the wall by diffusion depends only on mu = pi D L / Q_tube (Gormley and Kennedy 1949): 1 - P = 1 - 0.819 e^(-3.657 mu) - 0.0975 e^(-22.3 mu) - 0.0325 e^(-57 mu) - 0.0154 e^(-107.6 mu) for mu above 0.009, and 2.56 mu^(2/3) - 1.2 mu - 0.177 mu^(4/3) below. Going down the tree the dose that survives generation z is multiplied by (1 - efficiency_z); the deposit in generation z is what is lost there.
Stochastic breath
Every breath sends 200 particles. Each one is assigned its landing level by random draw from the model, so the counted fraction follows the theory with sampling noise.
Simplifications
Only diffusion is modelled, which is the main mechanism below about 100-200 nm; sedimentation and impaction (important above 0.5 micrometres and in the mouth and throat), nasal filtering, airway asymmetry, breath-by-breath volume changes, hygroscopic growth, mucociliary clearance and the extra surface of alveoli are all left out. Treating the alveolar region as plain tubes under-represents its true capture. Because of this the model overestimates total deposition for particles of roughly 20 to 100 nm compared with detailed dosimetry models, which also account for the dead-end geometry of alveoli. This is a teaching model, not a dosimetry tool.
What to try
- Start at 30 nm and read how the dose splits between the conducting airways and the alveoli.
- Press 5 nm: almost everything is caught in the trachea and bronchi before it can go deeper.
- Press 100 nm and then 300 nm: particles diffuse so slowly that most are breathed out again.
- Change the flow from 15 to 6 L/min: slower air gives more time to diffuse and deposition rises.
- Watch the tube colours turn orange where the dose lands and compare with the bar chart by generation.