The science
The device
A semiconducting carbon nanotube bridges a source and a drain electrode and is gated through the liquid above it. Receptor molecules (for example antibodies) are attached to the tube wall. When a target molecule binds, its electric charge changes the potential at the tube and shifts the transistor threshold voltage, which changes the current at a fixed gate bias. The 3D scene is magnified and not to scale: a real tube is about 1 nm in radius.
Binding: Langmuir kinetics
Each receptor site is a two-state system. With target concentration C, association constant k_on and k_off = k_on K_d, the bound fraction obeys d(theta)/dt = k_on C (1 - theta) - k_off theta, so theta(t) = theta_eq (1 - e^(-t/tau)) with theta_eq = C / (C + K_d) and tau = 1 / (k_on C + k_off). The 48 sites are updated with the exact Markov transition probabilities, so the average follows this curve and the scatter around it is real stochastic noise. Real experiments take minutes, so a time-compression slider speeds the clock up.
Debye screening
Salt ions gather around charged surfaces and screen them. In water at room temperature the Debye length is lambda_D = sqrt(eps kT / (2 N_A e^2 I)) = 0.304 nm / sqrt(I [mol/L]): about 9.6 nm at 1 mM, 3 nm at 10 mM and under 1 nm at 150 mM (blood-like salt). A charge sheet at height d above the surface produces a potential (sigma lambda_D / (eps_r eps_0)) e^(-d/lambda_D) at the tube (linear Poisson-Boltzmann with an insulating wall). A large antibody whose charge sits several nanometres up is almost invisible in physiological saline, which is why many sensors use diluted buffer or short receptors.
From charge to current
The threshold shift is dV_th = sigma_eff lambda_D / (eps_r eps_0) with surface charge sigma = theta N q e. The drain current uses a square law I = (k'/2)(V_gs - V_th)^2. The detection limit is the concentration where the shift reaches an assumed 1 mV noise floor: C = K_d theta / (1 - theta) with theta = 1 mV / dV_th(full).
Simplifications
One tube with a single receptor species, uniform site spacing, the charge of each bound target placed at one height, linear (Debye-Hueckel) screening valid only for small potentials, a fixed receptor density, no mass-transport limit, no non-specific binding, no gate-leakage or Schottky-barrier physics and an idealised square-law transistor. Values are order-of-magnitude, not a calibrated device model.
What to try
- Start with the default dilute buffer and watch the bound fraction rise toward its Langmuir equilibrium while the meter shows the threshold shift.
- Choose "Physiological saline": the Debye length collapses to under 1 nm and the shift almost vanishes even though just as many targets are bound.
- Then pick "Short receptors in saline": bringing the charge closer to the tube restores part of the signal.
- Lower the concentration to 1 nM with a strong binder and compare the equilibrium coverage with the detection limit readout.
- Open the side view camera and match the cyan Debye marker with the orange charge-height marker on the ruler.