Readouts
Presets
Generator and schedule
Independent numerical checks
The science
A Mo-99/Tc-99m generator is a decay chain. Mo-99 sits on an alumina column; its daughter Tc-99m is washed off with saline ("milked") and the column then regrows it.
A1' = -L1 A1 A2' = BR L2 A1 - L2 A2 (activities, Bateman). A2(t) = A2(0) e-L2 t + BR A1(0) L2/(L2-L1) (e-L1 t - e-L2 t). Peak ingrowth from an empty column at tmax = ln(L2/L1)/(L2-L1); later A2/A1 → BR L2/(L2-L1) (transient equilibrium). L = ln2 / half-life; half-lives used: Mo-99 65.94 h, Tc-99m 6.01 h; BR = 0.875.
In the 2D view the log graph makes the transient equilibrium visible: after a few Tc half-lives the Tc-99m curve is parallel to the Mo-99 curve (slope -L1), and each elution drops the Tc line back down so it regrows along the same ingrowth curve.
Simplifications, stated honestly. Half-lives and the 0.875 branching fraction are typical tabulated values quoted from memory and need checking against a nuclear-data table. The model ignores the Tc-99 daughter, Mo breakthrough, column chemistry, dead volume and radiolytic losses; elution yield is a constant; activities in GBq are illustrative. The bench is not to scale. This page teaches decay physics only; it gives no handling, dosing or clinical guidance, and makes no healing claim.
Cutaway sprite: Blender (Cycles) render of the 3D page's generator model.