Impedance Biosensor: Nyquist Plot and Electrode Surface (2D)

Analyte binds the surface probes, charge transfer gets harder and the Nyquist semicircle grows: read the signal, the apex frequency and the calibration curve.

Switch to 3D
Controls

Presets

Parameters

Model checks

The science

Impedance as a sensor signal. In electrochemical impedance spectroscopy a small AC voltage is applied at many frequencies and the complex impedance Z(ω) is recorded. For a label-free biosensor a redox probe (ferri/ferrocyanide) is added; when the target binds the surface probes, it blocks electron transfer and the charge-transfer resistance Rct rises. The signal is the change of Rct, read off the diameter of the semicircle in the Nyquist plot.

Randles circuit. Z = Rs + (Rct + ZW) / (1 + jωCdl(Rct + ZW)), with the Warburg element ZW = σω-1/2(1 - j) for diffusion of the redox probe. At high frequency Z → Rs; without diffusion the low-frequency end is Rs + Rct and the semicircle has its apex at ω = 1/(RctCdl) where Z = Rs + Rct/2 - jRct/2. With diffusion the plot continues as a 45° line: -Z'' = (Z' - Rs - Rct) + 2σ2Cdl.

Binding. Probe-analyte binding is Langmuir: dθ/dt = konc(1 - θ) - koffθ, equilibrium θ = c/(c + Kd), time constant 1/(konc + koff), and Rct = Rct0 + ΔRmaxθ. The calibration curve is therefore hyperbolic with half-saturation at Kd; a concentration can be read back as c = Kdθ/(1 - θ).

Simplifications. Ideal Randles circuit (real electrodes need a constant-phase element instead of a capacitor), a linear dependence of Rct on coverage, a single binding site type, no non-specific adsorption and no drift (the biofouling and drift of implantable sensors is a different effect). Parameter values are typical for a screen-printed gold electrode, not measurements. Checks: the two frequency limits, the exact semicircle apex, impedance form against admittance form, the Warburg line, RK4 binding kinetics against the closed form, and the calibration inverse.

-
-