Windkessel Lab: Arterial Compliance, Pulse Pressure and Peripheral Resistance

A pump, a flexible chamber and a tap built in Blender model the arterial Windkessel effect: stiffen the arteries or close the tap and watch the pressure gauge and the trace. Drag to orbit.

Switch to 2D
Controls

Presets

Parameters

Quality

Camera

Model checks

The science

The idea. The heart ejects blood in pulses, yet the blood in the small vessels flows almost steadily. The large arteries are elastic: during ejection they swell and store volume, between beats they recoil and keep pushing blood on. This is the Windkessel (“air chamber”) effect. The simplest model is two elements: a compliance C (mL per mmHg) and a peripheral resistance R (mmHg·s per mL), C dP/dt = Q(t) − P/R, where P is the pressure above venous pressure and Q the inflow from the heart.

Exact solution. Ejection is a half-sine of duration Ts with stroke volume SV, Q = Qm sin(πt/Ts), Qm = π SV/(2 Ts). With τ = RC the pressure in systole is P = P0e−t/τ + a[sin ωt − ωτ cos ωt + ωτ e−t/τ], and in diastole it decays as e−(t−Ts)/τ. Requiring the beat to repeat gives the end-diastolic pressure in closed form. Two results are worth remembering: the mean pressure is always CO × R whatever the compliance, but the pulse pressure grows as the arteries stiffen (C falls); a short time constant makes the pressure fall far between beats, a long one makes it nearly flat.

Numerical checks. (1) The closed-form steady state is compared with RK4 after 120 beats; (2) the mean of the simulated cycle equals CO × R; (3) the least-squares slope of ln P in diastole gives the time constant RC; (4) the ejected volume equals SV and the volume stored per cycle is zero (balance of inflow and runoff); (5) the beat solution is compared with RK4 from a different starting pressure at five times.

Simplifications. A two-element model has no wave reflection, no characteristic impedance of the aortic root and one lumped artery; real compliance falls with pressure and age, ejection is not a perfect half-sine and venous pressure is not exactly zero. The numbers are textbook-style; this is an educational physiology model, not a diagnostic tool or medical advice.