The science
A PACS viewer shows numbers, not pictures: every pixel is a measured quantity and the viewer maps it to gray. This phantom is a synthetic thorax and abdomen (no patient data) built from analytic shapes, with tissue properties from published ranges.
CT. X-rays obey Beer-Lambert attenuation, I = I0 exp(-integral of mu ds). Reconstruction returns mu, expressed in Hounsfield units:
HU = 1000 (mu - mu_water) / mu_water (air -1000, water 0, bone > 300)
Image noise comes from photon statistics: SD scales as 1/sqrt(mAs) and 1/sqrt(slice thickness). Averaging thicker slices also blurs small lesions through partial volume.
Windowing. Screens show about 256 gray levels, CT holds thousands. The DICOM linear VOI LUT (PS3.3 C.11.2.1.2) maps window centre c and width w to gray y:
y = ((x - (c - 0.5)) / (w - 1) + 0.5) (ymax - ymin) + ymin, clipped to [ymin, ymax]
A narrow window (liver 60/150) shows a 22 HU lesion in 60 HU liver; a wide one (lung -600/1500) shows vessels in air. Same data, different question.
Stored values. CT pixels are stored as unsigned 12-bit integers; the header holds RescaleSlope m and RescaleIntercept b, so HU = m x stored + b (here b = -1024: air 0, water 1024). MPR resamples the same voxel grid on three orthogonal planes, using the pixel spacing and ImageOrientationPatient to keep millimetres true on screen.
MRI. Spin-echo signal depends on proton density, T1, T2 and the timing you choose: S = PD (1 - exp(-TR/T1)) exp(-TE/T2). Short TR and TE give T1-weighting (fat bright, fluid dark); long TR and TE give T2-weighting (fluid bright). Magnitude noise is Rician. Tissue values are typical 1.5 T figures.
Radiograph. A projection sums mu along every ray; the page computes a digitally reconstructed radiograph with quantum noise from the photon count.
Ultrasound. The probe sends pulses; echo amplitude at an interface is set by acoustic impedance, R = ((Z2 - Z1)/(Z2 + Z1))^2, and depth by time of flight, d = c t / 2 with c = 1540 m/s. Attenuation (dB/cm/MHz x frequency x two-way distance) is compensated by time-gain control; speckle is a correlated Rayleigh field. Air and bone almost fully reflect, so ribs and lung cast shadows.
PACS and DICOM. Scanner, router, archive and workstation talk DICOM over TCP: an association is negotiated, each image goes as a C-STORE of 16 KiB P-DATA PDUs and is acknowledged by a response. Time is about 2.5 RTT + N (size / throughput + RTT) + RTT, with throughput = min(link, 64 KiB window / RTT), so a long-distance link is limited by latency, not bandwidth; lossless JPEG 2000 reduces size by roughly 2-3x.
Checks. (1) The DICOM window formula at its defining points. (2) Measured CT noise SD against 14 sqrt(100/mAs) sqrt(4/thickness) and the 1/sqrt(dose) ratio. (3) Liver region mean against 60 HU. (4) Body width measured with the ruler against the analytic ellipse. (5) Beer-Lambert transmission through 100 mm of water. (6) Liver T1 recovered from two repetition times. (7) Reflection plus transmission equals one. (8) Echo time of a bone interface mapped back to depth. (9) Radon line integral of a disc against the analytic chord. (10) Filtered back-projection of a real slice from its sinogram. (11) DICOM transfer time: closed form against a PDU-by-PDU count.
Simplifications. Phantom tissue is piecewise uniform; CT uses one effective energy (no beam hardening or scatter); MRI ignores B1/B0 inhomogeneity and motion; ultrasound is a 2D ray model without refraction or multiple reflections; the network model has a single association and no packet loss. This is an educational tool, not for diagnosis.