Technetium-99m generator: molybdenum-99 ingrowth and elution

Switch to 3D
A fresh column holds only Mo-99. Wait about 23 h, then elute.

Readouts

Time-
Mo-99 in column-
Tc-99m in column-
Share of equilibrium value-
Eluate vials-
Peak ingrowth time-
The science

What is modelled. The parent-daughter decay chain inside a Mo-99 / Tc-99m generator: Mo-99 (T1/2 = 65.94 h) feeds Tc-99m (T1/2 = 6.0067 h) with branching b = 0.875; saline elution removes a fraction η of the daughter. This is an educational physics model, not clinical or dosing guidance.

Equations. dA1/dt = −λ1A1, dA2/dt = bλ2A1 − λ2A2. Solution from an empty column: A2 = bλ2/(λ2−λ1) A1(0)(e−λ1t − e−λ2t), peak at tmax = ln(λ2/λ1)/(λ2−λ1) = 22.8 h, then transient equilibrium A2/A1 = bλ2/(λ2−λ1) = 0.963 (dashed curve). The page steps the exact solution; the checks below compare it with RK4 and a stochastic simulation.

Honest simplifications. One fixed elution yield; the longer-lived Tc-99 that builds up is not tracked; Mo-99 activities (2 to 60 GBq) are illustrative; no breakthrough or purity limits; the picture is a Blender render of a schematic generator, not a specific product.

Independent checks (same core as the 3D page):

running…
Daughter-to-parent activity ratio A(Tc-99m) / A(Mo-99): rises to the transient-equilibrium value (dashed)
Eluate vials: each decays with the 6.0067 h half-life (usable window shrinks fast)
Made with 🟠 Blender