How to use
Choose a preset or move any slider. The picture shows the left ventricle (size follows volume), the aortic valve, the aorta coloured by pressure, the balloon in the descending aorta and the helium console; the strip chart shows the ECG trigger, the aortic and LV pressures and the balloon-inflated band, with dashed lines for the pump-off pressures. The graphs show pressures, the pressure-volume loop against the pump-off loop, the coronary driving pressure, and two experiments: how the inflation timing changes the pressures, and how the balloon volume changes the coronary driving pressure. Space pauses, R restarts. Try: (1) Shock, pump off, then Shock + IABP 1:1: the diastolic hump appears and the end-diastolic pressure falls. (2) Early inflation and Late inflation: watch the hump and the stroke volume. (3) Assist 1:2: compare the assisted and unassisted beats in the readout. (4) Healthy heart + IABP: a normal heart gains little.
The science
This is an educational lumped-parameter model of a failing left ventricle with an intra-aortic balloon pump (IABP). It uses synthetic round-number values and is not a clinical tool.
What the pump does. A long balloon sits in the descending aorta on a catheter. Triggered by the ECG, a console shuttles helium in and out: the balloon inflates just after the aortic valve closes (the dicrotic notch) and deflates just before the next beat. Inflation pushes blood back toward the aortic root and the coronary arteries (diastolic augmentation); the sudden deflation empties space in the aorta, so the ventricle ejects into a lower pressure (afterload reduction). This is counterpulsation.
The ventricle is a time-varying elastance (Suga and Sagawa; Stergiopulos double-Hill activation):
Plv(t) = E(t) (Vlv − V0), E(t) = Emin + (Emax − Emin) e(t/Tmax)
The aortic and mitral valves are ideal diodes with resistances. The arteries are a Windkessel (characteristic impedance Zc, compliance Ca, systemic resistance Rp) feeding a venous reservoir Cv.
The balloon enters as a displaced volume. If Vb(t) is the balloon volume, the blood it pushes out raises the arterial compartment pressure:
Ca dPa/dt = Qao − (Pa − Pv)/Rp + dVb/dt
so that Vlv + CaPa + CvPv − Vb is exactly conserved (check 1). A fast inflation of Vb into the compliance Ca raises the pressure by about ΔP ≈ Vb/Ca (check 3), less the blood that runs off through Rp and the valve during the 50 ms ramp. A stiffer aorta (small Ca) therefore gives a larger augmentation.
Timing. Inflation starts at the detected aortic valve closure plus your offset; deflation finishes a chosen lead time before the next R wave. Too early an inflation squeezes the still-ejecting ventricle (lower stroke volume); too late loses the first part of diastole; deflating too early lets the aorta refill before the beat, so the unloading (and the DPTI/TTI ratio) is weaker. The deflation lead slider starts at 0 ms, a deflation that just finishes as the next beat begins.
Indices. The tension-time index TTI = ∫Plvdt over systole is a proxy for oxygen demand; the diastolic pressure-time index DPTI = ∫(Pao − Plv)dt over diastole is a proxy for coronary oxygen supply (the coronary arteries fill mainly in diastole, driven by Pao − Plv). Their ratio DPTI/TTI is the Buckberg endocardial viability ratio; values below about 0.7 to 0.8 mean supply is short of demand. The pump raises DPTI and lowers TTI.
Numerical checks run in the page. (1) Volume conservation with the pump cycling; (2) mean(Pa) − mean(Pv) = CO Rp with the pump on; (3) the inflation pressure jump against Vb/Ca; (4) the balloon volume integral over one beat is zero; (5) the balloon peak equals the setting; (6) diastolic augmentation, (7) afterload reduction and (8) a higher DPTI/TTI in the shock state.
Honest simplifications. One ventricle; no right heart; the venous return is a single reservoir; the balloon is a volume source with a fixed 50 ms smooth ramp (real balloon dynamics depend on catheter length, helium gas dynamics and heart rate); no coronary flow model (the coronary driving pressure is a proxy), no autoregulation, no reflected waves (so the dicrotic notch is mild); the ECG is stylised; the values are plausible round numbers, not patient data. The pump supports a heart in cardiogenic shock but does not create flow by itself: the cardiac output gain is modest.