The science
What is modelled
Each step slides the femoral head across the liner under load and wears off a tiny volume, proportional to the load and the sliding distance. Over millions of steps the head sinks into the liner and sheds debris. The immune cells that eat the debris can also eat the bone around the implant; if too much bone is lost the implant loosens.
Equations
- Archard law
V = k ∫ F ds; for a half-sine load and a matching sliding sweep one step gives∫ F ds = Fpk r Δθ π / 4. - Linear penetration
h = V / (π r2), wear ratew = k N Fpk (Δθ π / 4) / π r. - Debris count
Np = V / (π d3 / 6). - Activity declining at a % per year gives
h(t) = w0 (1 − e−at) / a. - Bone support (a simple illustrative law)
dB/dt = −κ max(0, w − wth) Bwithwth = 0.05 mm/yrandκ = 0.4.
Checks run in the page
- Simpson integration of the stance work against the closed form.
- Grid quadrature of the wear volume against pi r^2 h.
- Monthly stepping of the wear depth against the closed form.
- Runge-Kutta bone loss against the closed form.
Honest simplifications
Real hips slide in several directions with changing load and lubrication, and the wear factor changes with time and material. The bone-loss law is a phenomenological illustration, not a clinical model, and the default numbers are typical orders of magnitude for a conventional polyethylene liner, not patient data. This page is for learning, not for medical decisions.
What to try
- Press Conventional polyethylene, then Cross-linked polyethylene and Ceramic on ceramic and compare the wear rate and the bone support.
- Press Large head, same material: the same material wears more with a bigger head.
- Press Heavy, very active patient and find the year the bone support crosses the loosening line.