Controls
Presets
Parameters
Model checks
The science
Force and speed. A rotor turning at angular velocity ω pushes a particle at radius r with the centrifugal acceleration ω2r. Laboratories quote it as the relative centrifugal force, RCF = ω2r/g = 1.118×10−5 r[cm] rpm2; at 3000 rpm and 10 cm that is about 1000 g. A single red cell (about 3.4 µm equivalent radius, 70 kg/m3 denser than plasma) reaches the Stokes velocity, so dr/dt = sω2r with s = 2a2Δρ/(9η) and r(t) = r0esω2t: the cell accelerates as it moves outward. Doubling the plasma viscosity halves s.
Crowded cells hinder each other. At normal hematocrit the plasma has to flow back through the cell cloud. The Richardson-Zaki law multiplies the speed by (1 − φ/φm)4.65, which stops sedimentation at the packing limit φm = 0.96 (some plasma is always trapped between cells). The page solves the conservation law φt + (φv)r = 0 in the tube with a Godunov finite-volume scheme; three zones appear: clear plasma next to the meniscus, a suspension, and a packed column against the tube bottom. The packed column has length H0L/φm, so the reading is slightly higher than the true hematocrit H0 because of trapped plasma. In a laboratory test the result is also read at a fixed speed and time, not by modelling.
Numerical checks. (1) The RCF definition agrees with the laboratory formula; (2) RK4 of the single-cell Stokes law crosses the tube in the analytic time; (3) cells are conserved by the closed-tube scheme and the interface of a dilute suspension follows the exponential law; (4) at the end of the spin the packed column equals H0L/φm; (5) the Stokes viscosity scaling is reproduced.
Simplifications. A one-dimensional constant-section tube, a single cell size, no deformation or aggregation kinetics (rouleaux are a single factor), no acceleration or braking of the rotor, no temperature effects and a fixed 5 cm sample column. This is an educational physics model of sedimentation, not a laboratory procedure.