The science
Serial dilution error
Each step transfers a volume Vt = Vw/DF into diluent Vd = Vw - Vt, so the concentration is multiplied by r = Vt/(Vt + Vd). A random volume error with coefficient of variation CV(V) = 0.003 + 0.12/V (V in uL, an illustrative air-displacement pipette shape) gives d ln r = (1 - r)(eT - eD). Errors add as independent random steps, so after k transfers SD(ln C) = sqrt(k) (1 - r) sqrt(CVT² + CVD²). Small volumes (large CV) and a re-used tip (carry-over of stronger solution) make it worse; a liquid handler is modelled as 0.35 times the manual CV.
Hill / four-parameter logistic
signal = bottom + (top - bottom) / (1 + (c / IC50)h). At c = IC50 the signal is exactly midway. The fit searches IC50 and h on nested grids; for every pair the linear parameters (top, bottom) follow in closed form by least squares.
Z' factor and Ki
Assay quality: Z' = 1 - 3(σpos + σneg)/|μpos - μneg| (Zhang, Chung and Oldenburg 1999); above 0.5 is considered excellent. For a competitive inhibitor Ki = IC50/(1 + [S]/Km) (Cheng-Prusoff) and ΔG = RT ln Kd with RT = 0.5925 kcal/mol at 25 C.
Checks
(1) A Monte-Carlo estimate of the SD of ln C after 7 dilution steps is compared with the closed form. (2) A noise-free dose-response is fitted and must return the exact IC50 and Hill slope. (3) Z', the free-energy conversion of a 1 uM binder and the Cheng-Prusoff relation are compared with their textbook values.
Simplifications
Plate-reader noise is Gaussian (proportional plus additive); edge effects, evaporation and compound solubility are ignored, and the pipette-error curve is illustrative, not a specific instrument's ISO 8655 limit. The model code is the same module the 3D lab uses.
What to try
- Pick "Rushed 3-fold, re-used tip" and watch the dilution error leave the +-1 SD band as the series goes on.
- Switch to the liquid handler and compare the spread of fitted IC50 over repeated runs (third graph).
- Shrink the well volume to 20 uL: the CV rises steeply (fourth graph).
- Raise reader noise until Z' drops below 0.5.