2D twin of the 3D deposition lab: material from the source lands on a vibrating quartz crystal; every added nanogram lowers its resonant frequency.

Presets

Playback

Deposition and crystal

Frequency shifts are tiny compared with 5 MHz but measurable to about a hertz, which is nanogram sensitivity.

Film

Model checks

The science

What is modelled

An AT-cut quartz disc vibrates in thickness-shear mode at 5 MHz. A thin, rigid, well-coupled film adds mass and lowers the frequency in proportion to its areal mass, which is how deposition rates are monitored in vacuum coaters.

Equations

  • Sauerbrey: Δf = -2 f0² (Δm/A) / √(ρq μq), sensitivity 56.6 Hz cm²/µg for 5 MHz.
  • Resonance curve: amplitude 1/√(1 + x²) with x = 2 δ/Γ, bandwidth Γ = f/Q.
  • Desorption (exact): h(t) = h0 e-kt + (r/k)(1 - e-kt).
  • Temperature: an illustrative AT-cut cubic drift term.

Checks run in the page

  • 100 nm of gold gives the Sauerbrey shift, and the sensitivity matches the textbook 56.6.
  • The full width of the resonance curve at 1/√2 equals f/Q.
  • A desorbing film follows exp(-kt) over 100 s.

Honest simplifications

Sauerbrey holds only for thin, rigid, uniform films; soft polymers, liquids and thick films need viscoelastic models (the polymer film here is treated as rigid). The Q-factor losses and the thermal term are empirical, not calibrated to a cut angle. The drawn film thickness is exaggerated.

What to try
  • Leave the shutter open with gold and read the frequency drop; close it and the frequency holds.
  • Raise the crystal temperature to 70 C and see the thermal part separate from the mass part on the first graph.
  • Choose Poor vacuum: the quality factor falls and the resonance curve widens.
Frequency shift against time
Film thickness against time
Resonance curve and the drive frequency
Sauerbrey line for each film