Vital Signs Monitor (3D)

A Blender bedside bay: the heart pumps, the pressure pulse runs down the artery and the monitor shows the same traces.

Presets

Playback

Heart

Blood vessels

Blood and breathing

Alarms and time

Camera

Layers

Quality

Model check

The science

What this page shows

A bedside monitor whose three traces come from a small physiological model, not from stored recordings: an ECG, the arterial blood pressure and the photoplethysmogram (pleth) of a finger pulse oximeter. Change the heart rate, stroke volume, vascular resistance and compliance, oxygen saturation or add noise, and the numbers and alarms respond. This is an educational model. It is not a medical device and must not be used to judge anyone's health.

Arterial pressure: two-element Windkessel

The arteries are treated as an elastic reservoir of compliance C (mL/mmHg) draining through the systemic resistance R (mmHg s/mL): C dP/dt = Q(t) - P/R. The heart ejects Q(t) = Qmax sin(πt/Tej) for the ejection time Tej (about 0.3 s at rest, shorter at high heart rate), with Qmax = π SV/(2 Tej) so the volume per beat is the stroke volume SV. Between ejections the pressure decays as exp(-t/RC). Averaged over a beat, dP/dt is zero, so mean pressure = cardiac output x resistance: MAP = SV x HR x R. Resistance is given in Wood units (mmHg per L/min; 1 WU = 80 dyn s cm-5). A stiff artery (small C) gives a larger pulse pressure.

ECG

The ECG is a sum of five Gaussians in the phase of the beat (P, Q, R, S and T), the algebraic form of the McSharry dynamical ECG model, scaled to an R wave of about 1.2 mV. Breathing modulates the beat interval (respiratory sinus arrhythmia) and adds baseline wander; the ST slider shifts the ST segment.

Pulse oximetry

The finger absorbs red (660 nm) and infrared (940 nm) light. Oxygenated and deoxygenated haemoglobin absorb differently (extinction coefficients used: HbO2 320 and 1214, Hb 3226 and 693 cm-1/M at 660 and 940 nm). Only the pulsing arterial blood changes with each beat, so the ratio of the pulsatile absorbances, R = ΔA660/ΔA940, depends on saturation alone. The page measures R on the red and infrared traces and inverts the Beer-Lambert relation to get SpO2. Commercial devices use an empirical calibration such as SpO2 ≈ 110 - 25 R; both are plotted.

Checks run in the page

1. Mean simulated pressure against CO x SVR. 2. Diastolic decay time constant fitted to the simulation against R C. 3. Mean R-R interval of R peaks found in the generated ECG against 60/HR. 4. SpO2 recovered from the generated red and infrared signals against the setting.

Honest simplifications

A two-element Windkessel has no dicrotic notch, no wave reflection and no pressure amplification toward the periphery; it is a single compartment with venous pressure taken as zero. The pleth is the delayed pressure pulse scaled to a perfusion index, ignoring tissue optics, venous blood and motion beyond the noise slider. Heart-rate regulation, baroreflex and gas exchange are not modelled: the sliders are independent.

Loading model...
Loading Blender model...
One steady beat: aortic pressure, inflow and the RC decay
Pulse oximeter calibration: R against SpO2
Red and infrared pulses at the finger
Trends over the last minute