A Coriolis meter drives a thin-walled tube pair into oscillation. Fluid moving through the vibrating tubes experiences a Coriolis acceleration; the resulting twist creates a small phase difference between upstream and downstream pickoffs.
ṁ = ρ Q
Mass flow is density multiplied by volumetric flow. Here Q is the adjustable process input.
FC = 2 ṁ vtube
The simplified transverse-force relation uses the peak tube velocity vtube = 2π f A.
Δφ = Kφ ṁ
The illustrative pickup calibration Kφ is 12 mrad per kg/s. A real instrument uses its own geometry-specific calibration and signal processing.
f = (1 / 2π) √[ k / (mtube + ρ Vint) ]
A single-mode spring–mass approximation shows how added fluid mass lowers the tube resonance frequency.
Q = ∫A u(r) dA, u(r)=2ū[1−(r/R)²]
The independent check numerically integrates a fully developed laminar parabolic velocity profile over the circular bore and compares it with the analytical Q.
Simplifications: constant calibration factor; one effective vibration mode; fixed tube stiffness and dry tube mass; incompressible Newtonian fluid; fully developed laminar profile for the numerical flow check; no thermal drift, pressure sensitivity, damping, two-phase flow, or cross-coupling. Values are textbook relations applied to an illustrative scaled design, not a calibration model for a commercial meter.