A UV–Vis spectrophotometer selects a narrow wavelength band, passes it through a sample, and measures transmitted radiant power at a detector.
ε(λ) = εmax · exp[−½((λ − λpeak)/σ)²]A(λ) = ε(λ) · l · c ; T = 10⁻ᴬ = I/I₀Eγ = hc/λ = 1239.84/λ(nm) eVTband = ∫G(λ) · 10⁻ᴬ(λ) dλ / ∫G(λ) dλIllustrative inputs: absorption peak 520 nm, peak molar absorptivity 18,000 L·mol⁻¹·cm⁻¹, Gaussian absorption width 42 nm, incident power 1.00 mW, and a Gaussian instrument bandpass set by the FWHM control.
Numerical check: bandpass transmission is integrated twice with composite Simpson quadrature (200 and 400 panels). The displayed difference is the absolute convergence error in percentage points.
Simplifications: ideal detector linearity, no stray light, no scattering, no reflection loss, a homogeneous non-fluorescent sample, and a single Gaussian absorption band. This is a teaching illustration, not a calibrated instrument simulation.