Carrier-gas velocity
u_out = Q_out / A = 4 Q_out / (π d²)
Outlet volumetric flow is converted to superficial linear velocity using the capillary cross-section. The model assumes a circular, unobstructed column.
Column efficiency
H = A_v + B / ū + C ū ; N = L / H
The van Deemter relation represents eddy-free longitudinal diffusion and mass-transfer broadening in a capillary column. A_v, B and C are illustrative teaching constants; ū is a pressure-adjusted mean velocity.
Retention and resolution
t_R = t_M (1 + k) ; R_s ≈ (√N / 4) · ((α − 1) / α) · (k / (1 + k))
The resolution estimate uses an equal-width Gaussian peak approximation, selectivity α = 1.035 and retention factor k = 2.0. It is not a prediction for a real analyte pair.
Pressure profile and numerical check
p(x) = p_in − (p_in − p_out) x / L
The carrier gas is treated as ideal and isothermal, with a linear pressure profile from the selected inlet pressure to a fixed 100 kPa outlet. A 1,000-interval Simpson integration of 1/u(x) is compared with the analytic pressure-profile hold-up time.
Simplifications
- No extra-column volume, adsorption, peak tailing or detector dynamics.
- Constant temperature and illustrative transport coefficients.
- Geometry is a scaled educational cutaway; the physical model is not a product specification.
Illustrative design, not affiliated with any real institution or manufacturer.