Quadrupole Mass Filter

2D twin of the QUADRA QMS-4 spectrometer: ion paths in the x-z plane, the rod end view, the (a, q) stability diagram and a simulated mass scan.

Presets

Playback

Filter settings

Tuning sets RF amplitude so that the chosen mass has q = 0.690; the DC/RF ratio sets how narrow the passband is (0.1678 would be the apex).

Mass scan

Model checks

The science

What is modelled

Four rods carry a potential Φ = (U + V cos ωt)(x² - y²)/(2 r0²). An ion of mass-to-charge ratio m/z obeys the Mathieu equation; only ions inside the first stability region of the (a, q) plane have bounded transverse motion and reach the exit. The ion integrator is the same RK4 module as on the 3D page.

Equations

  • a = 4 e U / (m r0² ω²), q = 2 e V / (m r0² ω²).
  • Stability of the monodromy matrix: |trace| < 2 for both x (a, q) and y (-a, -q).
  • Axial speed v = √(2 E / m) gives the flight time through the rods.

Checks run in the page

  • The stability apex found by Floquet theory (a = 0.2370, q = 0.706).
  • Tuning gives q = 0.690 and a/q = 2U/V.
  • Trajectory integration: the tuned mass passes, masses far from it are rejected.

Simplified: ideal hyperbolic field (no fringing), an idealised round-rod collision radius, a coarse ion count for the scan.

What to try
  • Raise the DC/RF ratio towards 0.1678 and run the scan: the peak narrows.
  • Retune to m/z 65 and see which source species still passes.
  • Lower the axial energy: ions spend more RF cycles inside and the rejection gets sharper.
Stability diagram with the four source masses
Simulated mass scan (transmission against m/z)
Ions passed and rejected per species
Mathieu q and stability against m/z